| Service
& Support - DTX-200 DexaCare®
|
 |
What
do I do when the computer gives the error message
"fatal error reading text files"?
This error can be caused by too little DOS memory
(minimum required is 580k) or by bad *.HOF and
*.TOF files. *.TOF and *.HOF files are Temporary
Operating Files and can be deleted without damage
to the system. These files are generated each
time a scan is run. If the program does not delete
them after its use, they can cause problems. To
delete the *.HOF and *.TOF files please do the
following:
From the DTX main menu press F10 to exit
to DOS
The computer should show C:\DTX
Now type the following:
DEL *.HOF
press ENTER
Then type:
DEL *.TOF
and press ENTER
Now type DTX
and press ENTER
To check the memory you can type at the C: prompt
the following and make sure you have enough memory
to run the program:
C:\mem (You should have at least 580k of
memory available).
This should clear everything out of the temporary
files and let the system operate normally.
If this did not solve your problem, please contact
your distributor or Osteometer
Technical Support.
|
| |
..................................................................
go to top
|
 |
What
should the Autoexec.bat and Config.sys files read?
The autoexec.bat:
If you are running Microsoft DOS this file can
be edited by typing at the C: prompt: edit
autoexec.bat
If you are running IBM PC DOS this file can be
edited by typing at the C: prompt:
e autoexec.bat
The file should read as follows:
1. C:\DOS\SMARTDRV.EXE /X
2. @ECHO OFF
3. PROMPT $p$g
4. PATH C:\DOS
5. SET TEMP=C:\DOS
6. MODE CON CODEPAGE
PREPARE=((850) C:\DOS\EGA.CPI)
7. MODE CON CODEPAGE SELECT=850
8. KEYB DK,,C:\DOS\KEYBOARD.SYS
9. cd c:\DTX
10. DTX
The lines in yellow might be different in your
setup and that is OK.
Lines 7-9 might be different due there monitor
and keyboard setups. This is OK. Lines 9 and 10
will automatically start the DTX-200 DexaCare®
program when the computer is started.
The Config.sys:
This file can be edited by typing at the c: prompt
the following:
C:\ edit config.sys
The file should read as follows:
1. DEVICE=C:\DOS\SETVER.EXE
2. DEVICE=C:\DOS\HIMEM.SYS
3. DOS=HIGH
4. COUNTRY=045,,C:\DOS\COUNTRY.SYS
5. DEVICE=C:\DOS\DISPLAY.SYS CON=(EGA,,1)
6. FILES=30
The yellow lines do not affect the performance
of the software. They have to do with the country
and monitor setups.
If this did not solve your problem, please contact
your distributor or Osteometer
Technical Support.
|
| |
..................................................................
go to top
|
 |
How do you make a calibration disk after performing
an Accuracy Calibration?
Instructions to make a new calibration disk or
update the existing disk after ACCURACY CALIBRATION.
Insert a blank disk labeled CALIBRATION DATA FOR
SN **** in drive A. If updating, insert the calibration
disk for that scanner. You will need to move the
write protect tab to the unprotected position
(normally identified with an open lock symbol)
Looking at the back of the disk, it is in the
upper left hand corner. Up is write protect, down
is unlocked. Be sure to put it back in the up
position after the files are updated.
From the main menu of the DTX program press F10
to exit to DOS
At the C:\DTX> type COPY D2BEAMHD.DAT A:
press ENTER
You should get a response that 1 file copied.
(if you are updating the calibration disk, you
will be asked if its is OK to overwrite the existing
file. Answer Y)
At the C:\DTX> type COPY DXHARD.SYS A:
press ENTER
You should get a response that 1 file copied.
(if you are updating the calibration disk, you
will be asked if its is OK to overwrite the existing
file. Answer Y)
To verify that the files copied correctly.
Type A: and press ENTER
Type DIR and press ENTER
You should get a message that the disk contains
2 files
D2BEAMHD.DAT
DXHARD.SYS
Type C: and press ENTER and then type DTX
and press ENTER to return to the DTX program.
Place this disk with the program disks in the
back of the Operators manual.
If this did not solve your problem, please contact
your distributor or Osteometer
Technical Support.
|
| |
..................................................................
go to top
|
 |
How do I UPGRADE my software?
Instructions for UPGRADE of the DTX-200 DEXACARE®
software.
Insert disk 1:2 into drive A
From the DTX Main Menu press F10 exit to
DOS
Type CD\ and press ENTER
At the C:\ type A: and press ENTER
At the A:\ type UPGRADE and press ENTER
You will now get a DTX-200 DEXACARE® screen
that says confirm path. Press ENTER
It will now start reading the disk and upgrading
the software.
It will stop and ask if you want to upgrade the
*.PRN files. Answer YES by pressing ENTER
It will stop and ask about the Reference Dataset.
Answer YES by pressing ENTER
It will stop and say Please insert next disk…
Insert Disk 2:2 and press ENTER
After disk 2 is read it will ask again to insert
next disk or press Esc. Insert next disk. Continue
until all disks have been installed.
Press the Esc key.
This will take you to C:\DTX. Now type DTX
and press ENTER to restart the program.
Remove the disk from Drive A and put the new disks
in the Operators Manual. Be sure to keep the Calibration
disk with the new program disks.
If this did not solve your problem, please contact
your distributor or Osteometer
Technical Support.
|
| |
..................................................................
go to top
|
 |
How do you remove unwanted
files in the print manager?
To remove unwanted files in the print manager.
Please do the following:
From the DTX main menu press F10 exit to
DOS
The computer should show C:\DTX now type CD
DXPRINT and press ENTER
The screen should now show C:\>DTX\DXPRINT
Now type DEL *.* and press ENTER
It will tell you that the files will be deleted,
are you sure?
Answer Y and press ENTER
Now type CD\DTX and press ENTER
Now type DTX and press ENTER
This should clear everything out of the print
manager and allow you to print normally.
If this did not solve your problem, please contact
your distributor or Osteometer
Technical Support.
|
| |
..................................................................