Osteometer
MediTech, Inc.
12515 Chadron Ave.
Hawthorne, CA 90250, USA.
Tel: +1 310 978 3073
Toll free: 866 421 7762
Fax: +1 310 676 0948
E-mail: info@osteometer.com
|
|
|
|
|
|
|
|
X-ray
System:
|
55
kV, 300 µA |
| Energy
Separation: |
K-Edge
filtration |
| Detection
System: |
Solid
state dual energy sandwich detector |
| Calibration
System: |
Line
by line internal reference calibration |
| Scanning
Method |
Rectilinear
(100 mm x 100 mm) |
| Scan
Resolution: |
0.4
mm x 0.4 mm |
| Scanning
Sites: |
new
Region of Interest (nROI):
The ultra-distal sites of the radius and ulna. Location:
Proximal from end-plate of radius and ulna, 65% trabecular
bone
8 mm Distal:
The distal site of radius and ulna. Location:8 mm radius/ulna
distance. Size: 24 mm in proximal direction, 13 % trabecular
bone
¼ Distal:
The site of radius and ulnaLocation: Mid-shaft ¼ armlength
proximal from ulna end-plate. Size: +/- 10 mm, mainly
cortical bone |
| Scan
Time: |
2
min., distal section of forearm |
| In
vivo Precision: |
>99% |
| In
vivo Accuracy: |
>97% |
| Patient
dose: |
Skin
dose on forearm: 0.2 mGy per scan |
| Effective
Dose: |
0.1 µSv
per scan |
| Leakage
Radiation: |
Less
than 5
µSv per hour measured 2" / 5 cm from sides of cabinet |
| Scatter
Radiation: |
Less
than 0.25
µSv per hour measured 3.3 feet / 1 m from top
of cabinet |
| External
Shielding: |
None
required |
|
Size & Weight: |
12½" (H), 26" (L), 15½" (W). 49 lbs.
32 cm (H), 66 cm(L), 40 cm (W). 22 kg. |
| Power
Requirements: |
100-120
/ 200-220 / 230-240 V~,
50-60 Hz, 125 VA |
| Power
Requirements: |
59-86°
F / 15-30° C |
| Minimum
PC Requirements: |
IBM PC compatible 486 DX2, 66 MHz or
fasterVGA display
4 MB RAM, 80 MB hard disk
3,5" (1.44 MB) floppy disk drive
1 Serial port (NS 16550AN recommended)
1 Parallel port Minimum |
| Printer
recommendations: |
HP
DeskJet 600C series or MS DOS 6.22 compatible equivalent,
e.g. HP 815C or HP 950C |
| ©
2004 Osteometer MediTech, Inc. All rights reserved. |
|
|
|