X-ray
System:
|
55
kV, 300 µA |
| Energy
Separation: |
K-Edge
filtration |
| Detection
System: |
Solid
state dual energy sandwich detector |
| Calibration
System: |
Line
by line internal reference calibration |
| Scanning
Method: |
Rectilinear
(100 mm x 100 mm) |
| Scan
Resolution: |
0.4
mm x 0.4 mm |
| Scanning
Sites: |
new
Region of Interest (nROI):
The ultra-distal sites of the radius and ulna Location:
Proximal from end-plate of radius, 65% trabecular bone
8 mm Distal: The distal site of radius and ulna
Location:8 mm radius/ulna distance Size: 24 mm in proximal
direction, 13 % trabecular bone
¼ Distal: The site of radius and ulna Location:
Mid-shaft ¼ arm length proximal from ulna end-plate Size:
+/- 10 mm, mainly cortical bone |
| Scan
Time: |
2
min., distal section of forearm |
| In
vivo Precision: |
In
vivo Precision > 99% |
| In
vivo Accuracy: |
In
vivo Accuracy > 97% |
| Patient
Dose: |
Skin
dose on forearm: 0.2 mGy per scan |
| Effective
Dose: |
0.1
µSv per scan |
| Leakage
Radiation: |
Less
than 5 µSv per hour measured 5 cm from sides of cabinet
|
| Scatter
Radiation: |
Less
than 0.25 µSv per hour measured 1 m from top of scanner |
| External
Shielding: |
None
required |
| Size
and Weight: |
31"(H),
23"(L), 11"(W). 114 lbs.
79 cm (H), 59 cm (L), 28 cm (W). 52 kg. |
| Power
Requirements: |
100/120/200/220/230-240
V~, 50-60 Hz, 125 VA |
| Room
operating Temp: |
15-30°C
or 59-86°F |
| Minimum
PC Requirements: |
IBM
PC compatible 486 DX2, 66 MHz or faster VGA display
4 MB RAM, 80 MB hard disk
3.5" (1.44 MB) floppy disk drive
1 Serial port (NS 16550AN recommended)
1 Parallel port |
| Minimum
Printer Requirements: |
HP
DeskJet 600 Series |
| ©