Search osteometer.com
Internet Search

Osteometer MediTech, Inc.
12515 Chadron Ave.
Hawthorne, CA 90250, USA.
Tel: +1 310 978 3073
Toll free: 866 421 7762
Fax: +1 310 676 0948
E-mail: info@osteometer.com

News & Events | FAQs | Home
Products & Services
 <<DTX-200 DexaCare®   < Technical Specifications
 

X-ray System:

55 kV, 300 µA
Energy Separation: K-Edge filtration
Detection System: Solid state dual energy sandwich detector
Calibration System: Line by line internal reference calibration
Scanning Method: Rectilinear (100 mm x 100 mm)
Scan Resolution: 0.4 mm x 0.4 mm
Scanning Sites: new Region of Interest (nROI): The ultra-distal sites of the radius and ulna Location: Proximal from end-plate of radius, 65% trabecular bone

8 mm Distal: The distal site of radius and ulna Location:8 mm radius/ulna distance Size: 24 mm in proximal direction, 13 % trabecular bone

¼ Distal: The site of radius and ulna Location: Mid-shaft ¼ arm length proximal from ulna end-plate Size: +/- 10 mm, mainly cortical bone
Scan Time: 2 min., distal section of forearm
In vivo Precision: In vivo Precision > 99%
In vivo Accuracy: In vivo Accuracy > 97%
Patient Dose: Skin dose on forearm: 0.2 mGy per scan
Effective Dose: 0.1 µSv per scan
Leakage Radiation: Less than 5 µSv per hour measured 5 cm from sides of cabinet
Scatter Radiation: Less than 0.25 µSv per hour measured 1 m from top of scanner
External Shielding: None required
Size and Weight: 31"(H), 23"(L), 11"(W). 114 lbs.
79 cm (H), 59 cm (L), 28 cm (W). 52 kg.
Power Requirements: 100/120/200/220/230-240 V~, 50-60 Hz, 125 VA
Room operating Temp: 15-30°C or 59-86°F
Minimum PC Requirements: IBM PC compatible 486 DX2, 66 MHz or faster VGA display
4 MB RAM, 80 MB hard disk
3.5" (1.44 MB) floppy disk drive
1 Serial port (NS 16550AN recommended)
1 Parallel port
Minimum Printer Requirements: HP DeskJet 600 Series
 
Articles/References Technical Sepcs. Demo Buy Now
 
 
 